Healthcare Technology Management (HTM) of mechanical ventilators by clinical engineers

Abstract

Mechanical ventilator failures expose patients to unacceptable risks, and maintaining mechanical ventilator safety is an important issue. We examined the usefulness of maintaining mechanical ventilators by clinical engineers (CEs) using a specialized calibrator. These evaluations and the ability to make in-house repairs proved useful for obviating the need to rent ventilators which, in turn, might prove faulty themselves. The CEs' involvement in maintaining mechanical ventilators is desirable, ensures prompt service, and, most importantly, enhances safe management of mechanical ventilators.

Letter to the Editor

It is important to ensure appropriate quality control of mechanical ventilatory support. Medical devices in the hospital such as mechanical ventilators and anesthesia machines have been managed by doctors and nurses throughout the world [1–3]. In Japan, doctors and nurses have performed the maintenance of mechanical ventilators in the past [4]. Faulty mechanical ventilators were repaired by the distributor. However, since the late 1990s, clinical engineers (CEs) who have a certificate of completion and have completed established course requirements now perform the maintenance of mechanical ventilators in hospitals. The aim of this letter was to investigate the effectiveness of maintaining mechanical ventilators by CEs using a specialized calibrator in hospitals.

In our hospital, 30 mechanical ventilators (5 models) were used. These were all evaluated using a specialized PTS-2000 calibrator (Puritan-Bennett, Mansfield, MA, USA) at the time of after-use inspections and periodic testing, calibration, or maintenance. For each ventilator, we evaluated the following parameters: inspiratory flow, tidal volume, oxygen concentration, peak inspiratory pressure (PIP), positive end-expiratory pressure (PEEP), respiratory rate, and inspiratory time. Mechanical ventilators that were out of the acceptable ranges were classified as faulty. Faulty mechanical ventilators were repaired by calibration and replacing sensors and/or circuit boards.

We systematically checked mechanical ventilators about 2,500 times using a PTS-2000 calibrator during the period from January 2004 to December 2010, and there were a total of 151 cases of suspected mechanical ventilator failure. The number of faulty ventilators was 90 for ventilatory volume, 39 for oxygen concentration, and 22 for malfunctions. The faulty ventilators needing valve and/or sensor replacement were repaired in-house by CEs. The rare, seriously malfunctioning ventilators (there were a total of 12 cases of circuit board failure) were sent to the distributor for repair. Regarding response times for valve and/or sensor replacement, in-house repairs were completed immediately. However, if a machine needed to be repaired to the distributor, it takes from 2 h to half a day, as there are no distribution centers in Yamagata prefecture.

The number of mechanical ventilators rented has increased each year (198, 246, 288, 268, 313, 391, and 573 from 2004 to 2010). However, minor problems during daily practice that can be handled on the spot have been reduced because of after-use maintenance of mechanical ventilators by CEs (62, 60, 49, 47, 25, 10, and 12 from 2004 to 2010). Major failures (which necessitated a change-out of the mechanical ventilator) have been reduced because of after-use maintenance of mechanical ventilators by CEs (17, 11, 5, 3, 0, 4, and 0 from 2004 to 2010). Until 2003, CEs only checked and fixed broken ventilators. Most of the parts that were exchanged due to ventilator failure were batteries and sensors, parts that are recommended for periodic replacement. In the present study, consumable parts like batteries and sensors, which deteriorate over time, were at fault in the majority of cases. Therefore, we have routinely checked consumable parts, such as batteries and sensors, by using the PTS-2000 calibrator during inspections. And we have actively conducted calibrations and replaced consumable parts. Preventive inspection by CEs has decreased faulty ventilators. Thus, the involvement of CEs has improved mechanical ventilator safety.

During the 1980s, general-purpose-type mechanical ventilators were in use. Their operating principle was simple, and mechanical ventilators at that time had few circuit boards and/or sensors [5]. These mechanical ventilators were rugged and did not break down easily. Operator error accounted for most of the sudden malfunctions of mechanical ventilators [6, 7]. However, over the last two decades, technological developments have led to significant improvements in the performance of mechanical ventilators [8–11]. Modern critical care mechanical ventilators have circuit structures with numerous CPU boards and sensors [12]. Therefore, these mechanical ventilators have more failure points that increase the likelihood of failure. Blanch reported that spending an immense amount of time and cost could secure the reliability of an apparatus, but it was hard to anticipate the deterioration of mechanical ventilators; their safety and reliability will change significantly as ventilators are used or age [13]. The use of the PTS-2000 calibration analyzer has contributed to the discovery of previously undetected malfunctions in the new generation of high-performance mechanical ventilators. In this way, evaluating ventilation and carrying out in-house repairs has proved to be effective for obviating the chance of renting faulty units. Clinical engineering has decreased medical device failure which exposes patients to potentially harmful risks.

The most important item should be the checking of mechanical ventilators by CEs. CEs are certainly specialists in medical devices, and their involvement in maintaining mechanical ventilators is logical for the hospital, prompt, and most importantly safe. CEs fill an essential role for the safe operation of mechanical ventilators, and, more importantly, the CEs technology provides safe maintenance for many hospitals.

References

  1. Chapkowski S, Pucilo NP, Cane RD, Shapiro BA: Impact on ventilator-check time of an in-circuit computerized respiratory monitoring system. Respir Care 1984, 29: 144-146.

    CAS  PubMed  Google Scholar 

  2. Hartle A, Anderson E, Bythell V, Gemmell L, Jones H, McIvor D, Pattinson A, Sim P, Walker I, Association of Anaesthetists of Great Britain and Ireland (AAGBI): Checking anaesthetic equipment 2012: Association of Anaesthetists of Great Britain and Ireland. Anaesthesia 2012, 67: 660-668.

    Article  CAS  PubMed  Google Scholar 

  3. Bourgain JL, Baguenard P, Puizillout JM, Ankri JC, Damia E, Billard V: The breakdown of anesthesia equipment survey. Ann Fr Anesth Reanim 1999, 18: 303-308. 10.1016/S0750-7658(99)80055-0

    Article  CAS  PubMed  Google Scholar 

  4. Kawamata M, Mayumi T, Miyabe M, Namiki A: An unusual cause of anesthetic ventilator malfunction. Masui 1994–1997, 1992: 41. (In Japanese with English)

    Google Scholar 

  5. Richard JCM, Kacmarek RM: ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers. Intensive Care Med 2009, 35: 1662-1663. 10.1007/s00134-009-1581-6

    Article  PubMed  Google Scholar 

  6. Blanch PB: Mechanical ventilator malfunctions: a descriptive and comparative study of 6 common ventilator brands. Respir Care 1999, 44: 1183-1192.

    Google Scholar 

  7. Vignaux L, Tassaux D, Jolliet P: Evaluation of the user-friendliness of seven new generation intensive care ventilators. Intensive Care Med 2009, 35: 1687-1691. 10.1007/s00134-009-1580-7

    Article  PubMed  Google Scholar 

  8. Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougère S, Harf A, Brochard L: Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 2002, 28: 1049-1057. 10.1007/s00134-002-1311-9

    Article  PubMed  Google Scholar 

  9. Tassaux D, Michotte JB, Gainnier M, Gratadour P, Fonseca S, Jolliet P: Expiratory trigger setting in pressure support ventilation: from mathematical model to bedside. Crit Care Med 2004, 32: 1844-1850. 10.1097/01.CCM.0000138561.11634.6F

    Article  PubMed  Google Scholar 

  10. Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L: A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 2009, 35: 1368-1376. 10.1007/s00134-009-1467-7

    Article  PubMed  PubMed Central  Google Scholar 

  11. Marchese AD, Chipman D, de la Oliva P, Kacmarek RM: Adult ICU ventilators to provide neonatal ventilation: a lung simulator study. Intensive Care Med 2009, 35: 631-638. 10.1007/s00134-008-1332-0

    Article  PubMed  Google Scholar 

  12. Chatburn RL: Computer control of mechanical ventilation. Respir Care 2004, 49: 507-517.

    PubMed  Google Scholar 

  13. Blanch PB: An evaluation of ventilator reliability: a multivariate, failure time analysis of 5 common ventilator brands. Respir Care 2001, 46: 789-797.

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jun Yoshioka.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

KK, MN, and JY conceived the study, participated in the design of the study and data interpretation, and drafted the manuscript. All authors read and approved the final manuscript.

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yoshioka, J., Nakane, M. & Kawamae, K. Healthcare Technology Management (HTM) of mechanical ventilators by clinical engineers. j intensive care 2, 27 (2014). https://doi.org/10.1186/2052-0492-2-27

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/2052-0492-2-27

Keywords